私密直播全婐app免费大渔直播,国产av成人无码免费视频,男女同房做爰全过程高潮,国产精品自产拍在线观看

Links
Contact Info.
  • Address:西安市高新區新型工業園信息大道17號
  • Zip:710119
  • Tel:029-88887506
  • Fax:029-88887506
  • Email:wangll@opt.ac.cn
Current Location :> Home > Publications > Text
Fluorescence hydrogen peroxide probe based on titanium dioxide film modified microstructured polymer optical fiber
writer:Dongdong Li, Lili Wang
keywords:MPOF
source:期刊
Issue time:2010年

A highly sensitive microstructured polymer optical fiber (MPOF) probe for hydrogen peroxide was made by forming a rhodamine 6G-doped titanium dioxide film on the side walls of array holes in an MPOF. It was found that hydrogen peroxide only has a response to the MPOF probe in a certain concentration of potassium iodide in sulfuric acid solution. The calibration graph of fluorescence intensity versus hydrogen peroxide concentration is linear in the range of 1.6 × 10−7 mol/L to 9.6 × 10−5 mol/L. The method, with high sensitivity and a wide linear range, has been applied to the determination of trace amounts of hydrogen peroxide in a few real samples, such as rain water and contact lens disinfectant, with satisfactory results.